Template: Change Evaluation List

Sven Piechottka
Updated April 26, 2024
0 comments

Template Download

This is a free template, provided by OpenRegulatory.

If you are a user of Formwork, our eQMS software, you can save a lot of time by choosing “QMS” on the top menu and “OpenRegulatory Templates” on the left menu, and then opening the relevant folder to find this template ready to load into Formwork.

If, for some mysterious reason, you’re using a different QMS Software, you can also simply download this template – specifically, as Word (.docx), PDF, Google Docs or Markdown file. Scroll down for a preview!

The template license applies (don’t remove the copyright at the bottom, don’t re-use this for commercial purposes).

Lost In Regulation? Book a Free 30-Minute Consulting Call.

Unsure how to get started and how to get your EU MDR medical device certified?
We’ve already helped hundreds of companies with their MDR compliance.
Book a free 30-minute consulting call and let’s discuss how you can get your compliance done efficienty.

Book a free 30-minute consulting call
Template preview
Regulation / GuidanceDocument Section
Medical Device Regulation, Annex IX
Chapter II Section 4.10
All
MDCG Guidance Document 2020-03All

Summary:

This list is used to document the evaluation of all regular change requests according to the company’s change management process.

Evaluation of Product Changes

Disclaimer for Use #1: You may want to separate your lists for product and organizational changes, as you should release a new list for every product version. This is to separately identify all changes related to that specific version. Each product version list should ideally be attached to the respective technical documentation file.

Disclaimer for Use #2: If any of the below mentioned categories are answered with YES, this indicates that your change must be classified as significant.

Disclaimer for Use #3: In the example content filled into the table below, “PCR” stands for “Product Change Request” and “OCR” stands for “Organization Change Request”.

Please note:

  • YES in the first two categories (intended use / essential requirements / GSPR) always leads to a significant change.
  • If the device is modified (a) to correct an error, for which there is a safety risk to the patient if the error is not corrected or (b) as part of field safety corrective actions for an incident, the change is discussed with the Notified Body to determine the significance of the change. If no Notified Body was involved in the conformity assessment process of the device, the change is treated as a significant change.
Evaluation Categories
Change Request ID#PCR01#PCR02(…)
Change Request DescriptionPerformance update of machine learning algorithm modelAdditional integrability of device with new clinical IT-system
Overall Evaluation OutcomeNot significantSignificant
Does the change alter specifications provided in the Intended Use? (Chart A of MDCG 2020-03, e.g. new user or patient population)NoNo
Does the change affect conformity with the General Safety and Performance Requirements?NoNo
Does change implementation require further clinical or usability data to support safety and performance? (Chart B of MDCG 2020-03)No, but internal validation showed that performance on same metrics improvedYes
Do new risks require risk control measures or are existing risks negatively affected? (Chart B of MDCG 2020-03)NoYes
Does the change alter built-in control mechanisms or alarms? (Chart B of MDCG 2020-03)NoNo
Does the change modify an operating principle or sources of energy? (Chart B of MDCG 2020-03)NoNo
Does the change introduce a new or major change of the operating system or of any component? (Chart C of MDCG 2020-03, e.g. major SOUP update)NoNo
Does the change introduce a new or modified architecture or database structure, change of an algorithm? (Chart C of MDCG 2020-03, e.g. changes to prediction principle of an algorithm model)No, same prediction modelNo
Does the change replace previously required user input by a closed-loop algorithm? (Chart C of MDCG 2020-03)NoNo
Does the change introduce a new diagnostic or therapeutic feature, or new channel of interoperability?
(Chart C of MDCG 2020-03)
NoYes, new channel of interoperability
Does the change introduce a new user interface or presentation of data? (Chart C of MDCG 2020-03)NoNo
Does change implementation involve changes in critical suppliers? (Chart D of MDCG 2020-03)NoNo
Does the change impact the way medical data is read or interpreted by the user, such that the treatment or diagnosis of the patient may be altered when compared to the previous version of the software?Yes, but only improved accuracy in supported diagnosisNo

Evaluation of Organizational Changes

Please note: YES in the first two categories (Intended Use / GSPR) always leads to a significant change.

Evaluation Categories
Change Request ID#OCR1#OCR2(…)
Change Request DescriptionChange of Company Business AddressAdding a new compliance process to comply with anti-bribery provisions
Overall Evaluation OutcomeSignificantNot significant
Does the change affect QMS conformity with the General Safety and Performance Requirements (MDR)?NoNo
Does the change affect product conformity with the General Safety and Performance Requirements (MDR) or the approved type / design?NoNo
Does the change relate to manufacturing processes, technologies, facility or equipment in a way that could impact product safety and performance?NoNo
Does the change affect the location of the company’s activities?YesNo
Does change implementation involve changes in critical suppliers?
(Chart D of MDCG 2020-03)
NoNo
Does the change affect the arrangements implemented to achieve continued compliance of the QMS with the relevant harmonized standards and/or MDR requirements (e.g. design verification, design validation, organizational structure, process interaction, quality control procedures)?NoNo

Based on the linked guidance documents, examples for non-significant changes are:

  • a software change that only introduces non-therapeutic and/or non-diagnostic features such as printing, faxing, improved image clarity, reporting format or additional language support
  • a software change that only modifies the appearance of the user interface with negligible risk of impacting the diagnosis or therapy delivered to the patient
  • a software change only intended to correct an inadvertent logic error that does not pose a safety risk and brings the system back into specification
  • a software change that consists only of tightening of design specifications within specified tolerances and where there is no creation of new features
  • changes to labelling to include additional languages required in other regulatory jurisdictions
  • minor changes to clarify the existing wording of warnings and precautions. However, in the case where these changes add or remove a contraindication, or remove a warning or precaution, the Notified Body shall be involved.

Based on the linked guidance documents, examples for significant changes are:

  • an alteration in software that modifies an algorithm impacting the diagnosis or the therapy delivered
  • introduction or removal of a new alarm function from the software such that a response to the new configuration may change the treatment of the patient in comparison to the previous version of the software
  • medical data is presented in a new format, new dimension or new measuring unit.

Template Copyright openregulatory.com. See template license.

Please don’t remove this notice even if you’ve modified contents of this template.

Template preview

Comments

Leave the first comment