Regulatory Requirement | Document Section |
---|---|
MDR Annex I (General Safety and Performance Requirements), Annex IV (EU Declaration of Conformity), Art. 27, 29, 31 (UDI / Eudamed) | All |
MPDG §8, §96, §97 | All |
ISO 13485:2015, Sections 4.2.3 | All |
Summary
This SOP describes how a medical device is brought to the market while observing applicable regulatory requirements, including the compilation of a Technical Documentation and issuing a EU Declaration of Conformity for the medical device.
Process Owner | <enter role of process owner> |
Key Performance Indicators | <enter KPIs to be tracked for the Management Review> |
Process Steps
1. Decision on New or Revised Medical Device
This process is initiated once the Management decides to place a new or updated medical device on the market. Business input could be:
- Conversations with prospective customers
- Market opportunities
- Internal ideas
If applicable, a responsible Product Manager for a new device is appointed. Together with the QMO, they determine the classification of a new device as part of the MDR Classification document.
Participants | Management, Product Manager, QMO |
Input | Management decision on new device |
Output | Completed medical device classification |
2. Application for Conformity Assessment by Notified Body
If necessary based on the medical device classification, a Notified Body is involved for the conformity assessment for the new device.
NOTE: a Notified Body must be involved in the conformity assessment of all medical devices except for class I.
Participants | Product Manager |
Input | Medical device classification |
Output | Contract with Notified Body |
3. Compilation of Technical Documentation and Declaration of Conformity
The technical documentation for the medical device is compiled by following respective processes for development (e.g. software development, risk management, usability testing). The Product Manager is responsible to coordinate this process step with all relevant members of the organization. Documentation for different versions of the device is kept separate.
The Product Manager maintains a Medical Devices List according to MDR Annex II Sec 1.2 that provides an overview of:
- Installed product versions per customer site (if applicable)
- Previous device versions, incl. release and decommissioning date
- Similar devices
Prior to submission to the Notified Body for conformity assessment (if applicable), the QMO checks the documentation for completeness according to the stipulations listed below. Compliance with EU legislation is documented in the MDR General Safety and Performance Requirements Checklist.
- MDR General Safety and Performance Requirements (Annex I)
- MDR EU Declaration of Conformity (Annex IV)
After successful approval by the Notified Body, the Management signs the declaration of conformity. A product (version) is considered released and placed on the market with the release of its declaration of conformity.
Participants | Product Manager, QMO |
Input | Technical documentation |
Output | Approval by Notified Body Signed EU Declaration of Conformity Updated List of Medical Devices and Product Versions |
4. Registration of Medical Device (UDI)
Country-specific requirements for registration are observed before making a new medical device (version) available in a market. For the European market, the Product Manager follows the requirements set out in Art. 29 and Annex VI MDR by registering the device’s Unique Device Identifier (UDI) in the European Database for Medical Devices (Eudamed).
For the implementation of product changes, the Product Manager applies for a new UDI-DI if the changes entail one of the following:
- A significant change based on the Change Evaluation List
- Changed intended use
- New product features
- Changes in product interoperability, new interface or connected systems
- Non-trivial change to the user interface
- New supported operating system
- New programming language
- New algorithm model
- Non-trivial change to the software or database architecture
The Product Manager assigns a new UDI-PI for modifications such as:
- Bug fixes
- Security patches
- User interfaces (limited to usability and excluding security improvements)
The UDI-PI is determined by the manufacturer. Typically, the device version number serves as the UDI-PI (see SOP Software Development).
NOTE:
- The Basic UDI-DI (or: BUDI-DI) is assigned by UDI providers and identifies a group of medical devices with the same purpose, risk class, comparable design and manufacturing features.
- The UDI-DI is assigned by UDI providers for each medical device product type and is revised for any major modifications (substantial updates) to the device.
- The UDI-PI you can assign yourself as a manufacturer for every single product and is revised for minor modifications (non-substantial updates).
Example:
- Basic UDI-DI: Wrist-band health tracking device
- UDI-DI: different code based on color and packaging (white/blue/red; single-/double-pack)
- UDI-PI: different code based on series/batch/version, expiration date, etc.
NOTE: As of early 2021, the European Commission has designated four entities to issue UDI codes: GS1, HIBCC, ICCBBA and IFA. As long as Eudamed is not set up for registration, national registration requirements apply. For Germany, consider §96 and §97 of the MPDG. At OpenReg, we recommend obtaining a UDI already to be prepared for the regulatory transition.
NOTE: As part of the registration process, authorities in non-EU countries may require the documentation to be provided in other languages. Consider norms and regulations for the translation of medical device documentation.
Participants | Product Manager |
Input | Released Declaration of Conformity |
Output | Updated or new UDI Completed device registration |
Template Copyright openregulatory.com. See template license.
Please don’t remove this notice even if you’ve modified contents of this template.