This is a free template, provided by OpenRegulatory.
If you are a user of Formwork, our eQMS software, you can save a lot of time by choosing “QMS” on the top menu and “OpenRegulatory Templates” on the left menu, and then opening the relevant folder to find this template ready to load into Formwork.
If, for some mysterious reason, you’re using a different QMS Software, you can also simply download this template – specifically, as Word (.docx), PDF, Google Docs or Markdown file. Scroll down for a preview!
The template license applies (don’t remove the copyright at the bottom, don’t re-use this for commercial purposes).
Lost In Regulation? Book a Free 30-Minute Consulting Call.
Unsure how to get started and how to get your EU MDR medical device certified? We’ve already helped hundreds of companies with their MDR compliance. Book a free 30-minute consulting call and let’s discuss how you can get your compliance done efficienty.
This SOP describes how CAPAs are implemented and tracked.
Process Owner
<enter role of process owner>
Key Performance Indicators
<enter KPIs to be tracked for the Management Review>
Process Steps
1. Input for CAPA
Various events may lead to creation of CAPA. Examples include:
Product or QMS non-conformities
Customer complaints
Internal bug reports, e.g. by developers
Audit findings
Post-market surveillance findings, including trends
Management review findings, including trends
These inputs may be received from any person inside or outside the company. The QMO is responsible for creating the CAPA and tracking its resolution.
CAPAs are tracked in the CAPA list.
Participants
QMO
Input
Output
Non-conformity, complaint, etc.
CAPA created
2. Decision on Next Steps and Immediate Action
If immediate action is necessary (e.g. field safety corrective action or a notification to authorities according to SOP Vigilance), the QMO consults the Person Responsible for Regulatory Compliance. Immediate action is carried out without undue delay (see ISO 13485 para. 8.5.2).
In any case, the QMO discusses the next steps with the person closest to the issue, e.g. for software bugs, the Head of Software Development.
Participants
QMO
Medical Device Safety Officer / Person Responsible for Regulatory Compliance (optional)
Input
Output
CAPA
CAPA, updated with action
3. Root Cause Analysis
The QMO coordinates a root cause analysis with the person closest to the issue. The preferred method for this is Five Whys. The result is added to the CAPA list.
Participants
QMO
Other people in company (optional)
Input
Output
CAPA
CAPA, updated with root cause
4. Implementation of Action
The QMO coordinates defining and implementing corrective and preventive action. Additionally, the QMO takes into account adverse negative implications and verifies that the actions do not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device. Outcomes are documented in the CAPA list.
Participants
QMO
Other people in company (optional)
Input
Output
CAPA
CAPA, updated with action plan
5. Verification and Check of Effectiveness
The QMO conducts the verification and effectiveness review of the implemented action. These are defined as below. Thereafter, the QMO closes the CAPA.
Verification: Documenting proof of implementation of actions taken.
Effectiveness: Review of the effectiveness of actions taken.
Participants
QMO
Input
Output
CAPA
CAPA, updated with verification, effectiveness review, closed date